What you can do to prolong the life of your instruments
Dentalight Preventive Handpiece Maintenance Series
A handpiece may be used for only a few minutes during a procedure, but what happens afterward can influence how long it lasts between rebuilds.
In Part 1, Premature Handpiece Failure: How the Handpiece Is Used, we looked at workload, handpiece rotation, air pressure, cutting demands, and bur condition. In Part 2 of our three-part series on premature handpiece failure, the focus is on what happens between patients: lubrication, cleaning, sterilization, and air quality.
1. Lubrication Must Reach The Components That Need It
Most offices lubricate on a schedule. The important question is whether the lubricant reaches the parts that need it.
A handpiece can be lubricated regularly and still experience premature wear if the wrong port is used, or if you assume the lubricant is distributed through the handpiece as intended. For example, you might lubricate a low-speed motor and expect the lubricant to reach the contra angle and head attached to it, which it will not. We see many heads that are bone dry.
Highspeed turbines, lowspeed motors, contra angles, and electric attachments all have their own combinations of bearings, gears, and other moving components. Simply putting lubricant into an instrument doesn't guarantee it reaches the head.
2. Things To Watch For When Using Automatic Maintenance Machines
Automatic maintenance machines can streamline handpiece care operationally and reduce staff training time. They can also make it easier to process multiple instruments throughout the day in busy practices, but they need to be monitored.
Lubricant supplies can run low. The wrong program can be selected. The machine itself can also develop problems that affect lubricant delivery or purge performance, without being obvious.
Assuming every completed cycle means the handpiece was maintained exactly as intended can cause many instruments to fail before you know it. If several handpieces begin showing unusually short service life around the same time, the maintenance machine should be one factor you investigate.
In a future post, we will take a closer look at the advantages and disadvantages of automatic maintenance machines compared with aerosol and dropper bottle lubrication.
3. Cleaning, Lubrication and Sterilization Work As A System
Handpieces are built to withstand sterilization, but problems can develop before they enter the autoclave, such as:
-
Internal debris: Leftover contamination gets baked inside bearings during autoclaving.
-
Harsh chemicals: Using inappropriate cleaning and cold sterilant products that degrade materials.
-
Purging issues: Inadequate lubrication or excess oil left in the instrument.
Simplify Your Workflow
You can easily eliminate these risks by switching to a one-step synthetic cleaner and lubricant. This single change streamlines your practice in two major ways:
-
Eliminates human error: Staff no longer have to manage multiple products or remember a specific order. It prevents the common mistake of confusing a standalone cleaner for a lubricant.
-
Cleans the instrument and survives the autoclave: A properly formulated synthetic solution is engineered to remain stable under repeated high temperatures.
The Premium Synthetic Advantage
Not all lubricants are created equal. Mineral based oils, including some major name brands, oxidize and thicken when exposed to heat and moisture. This leaves a sticky residue inside the handpiece that causes mechanical drag, invites contamination, and degrades performance. A premium synthetic lubricant will not break down while maintaining both lower viscosity and superior protection.
4. The Hidden Impact Of Air Supply Quality
Air pressure is routinely checked in an operatory, but other air quality factors are just as important.
If your compressor’s condition, filtration, or drying systems are neglected, moisture, degraded compressor oil and particulates will enter the dental unit’s air lines and your instruments. With the turbine operating at hundreds of thousands of RPM, small amounts of contamination matter.
Air-quality problems can also be difficult to recognize because the handpiece may continue to operate normally for some time. Watch for patterns. If similar problems begin appearing across several handpieces, the air supply deserves scrutiny.
What Happens Between Patients Adds Up
Premature failure is usually not caused by one dramatic mistake. More often, small problems are repeated over hundreds of operating and sterilization cycles.
That is why handpiece maintenance is less about checking tasks off a list and more about controlling the conditions the instrument experiences over time.
In the final part of this series, we will look at another group of causes: burs, chucks, spindle problems, physical damage, and what failure patterns can tell you about a larger problem outside the handpiece itself.
Join the conversation. Do you have a problem or question about handpiece maintenance you would like addressed in this series? Ask Neal.